| TECHNICAL SPECIFICATIONS |
| Partial discharge measurement channels |
Integrated processing functions |
| Connector | Female BNC |
| Impedance | 50 Ω |
| Coupling | AC |
| Voltage ranges | From ±5 mV to ±25 V |
| Vertical resolution | 14 bits |
| Bandwidth | 0 - 30 MHz |
| Sampling rate | 100 MS/s |
| Memory Depth | 2 MS per channel |
| Surge protection | Discharger and disconnection |
| Analog filter |
- High pass of 30 kHz
- High pass of 1 MHz
|
| Dynamic Range | 81 dB |
| Minimum pulse rise time | 3.5 ns |
|
| DP filtering |
Detection of discharges under the noise level by means of Wavelet filtering + statistical analysis |
| PD pulse waveform analysis |
Extraction of f, α and β parameters, for 3D clustering |
| Reference voltage waveform analysis |
Extraction of the parameters RMS value, frequency and offset of the AC signal |
| Storage of capture parameters |
Trigger moment (system time and GPS time with optional module), trigger type, result of armed (correct/fail) |
| GPS status storage (optional module) |
UTC time, coverage values, positioning mode, antenna position, self-survey, etc… |
|
| Voltage reference measurement channel |
Communication network |
| Connector | Female BNC |
| Impedance | 1 MΩ |
| Coupling | AC |
| Voltage range | From ±220 mV to ±40 V |
| Vertical resolution | 12 bits |
| Bandwidth | 0 – 16 kHz |
| Sampling rate | 100 kS/s |
| Memory depth | 2 kS |
|
| Connector type | RJ-45 |
| Communication Protocol | IEEE 802.3 |
| Speed | 10/100/1000 Fast/Gigabit Ethernet |
| Type of cable | RJ-45 8 pin |
|
| Temperature sensors |
Synchronization of multiple SmartVisor equipment |
| Connector type | Terminal connection with 0.05-1.3 mm² (16-30 AWG) |
| Number of sensors | Up to 9 sensors |
| Supported sensors | NTC thermistor with 10 kΩ |
| Maximum length of connection cable | Up to 30 meters |
|
| Connector type | MicroUSB |
| Type of cable | 20 cm microUSB cable |
| Maximum number of units in a chain | Up to 9 units - 18 simultaneous PD channels |
|
| Temperature and humidity sensor |
VHF/UHF measurement with UHF-CO2 converter |
| Connector type | Terminal connection with 0.05-1.3 mm² (16-30 AWG) |
| Number of sensors | 1 sensor |
| Supported sensor | AM2315C |
| Communication interface | I2C |
| Maximum length of connection cable | Up to 1.5 meters |
|
| Supported PD channel |
- PD1: Using external VDC coupler
- PD2: Integrated coupler activatable in channel
|
| Type of sensors | TEV, Windowed, Invasive / Drain Valve |
| Bandwidth |
- TEV: VHF/UHF 30 MHz to 1 GHz, 250–750 MHz with band-pass filter for UHF band
- Windowed/Invasive: UHF 100 MHz – 2 GHz
|
| DC Output Voltage | 12 VDC (coupled in measurement) |
|
| Current sensors |
Leds indicators |
| Connector type | Terminal connection with 0.05-1.3 mm² (16-30 AWG) |
| Number of sensors | Up to 4 sensors |
| Sensor type | Rogowski coil |
| Inner diameter | Up to 94 cm |
| Nominal coil output rate | 100 mV / kA @ 50/60 Hz (RMS) |
| Maximum measurable current | 100 kA |
| Integrator output | 0 to 10 VDC |
| Total scale values |
Configurable in amps: 100, 200, 300, 400, 500, 600, 800, 1000, 1500, 2000, 2500, 4000, 5000 |
| Degree of protection | IP68 |
|
| Number of indicators | Two, status and diagnostic |
| LED Type | RGB |
| Color Range | Green/Yellow/Orange/Red |
| Indication type | Fixed and blinking light |
|
| Pressure sensors |
General system functions |
| Connector type | Terminal connection with 0.05-1.3 mm² (16-30 AWG) |
| Number of sensors | Up to 2 sensors |
| Sensor type | Transducer |
| Connection | 2-wire interface |
| Analog output | 4 to 20 mA |
| Additional power supply | 24 V |
|
| Auto recovery | Watchdog |
| Configuration storage | Non-volatile flash memory storage |
| Monitoring data storage | SD card with 32 GB |
| System time | Real time clock |
|
| Advanced trigger function |
Network functions |
| Trigger conditions |
- Manual trigger
- Per level (edge/level)
- PPS of GPS (optional module)
|
| Pre-trigger | 16000 samples of 10 ns |
| Post-trigger | Up to 232 samples of 10 ns |
|
| IP Configuration | Static or dynamic IP |
| Integrated network services |
- Detection of network equipment by UDP
- Reception of control connections
- Send data to remote server in autonomous mode
|
|
| — |
Firmware update interface |
| |
| Local programming | JTAG Connection |
| Remote programming | FOTA Remote Update App |
|
| — |
Grounding |
| |
| Grounding | By M3 screw on panel |
|
| — |
Power supply |
| |
| Connector type | Terminal connection with 0.05-3.3 mm² (12-30 AWG) |
| Voltage range | 12-24 VDC / 12 VAC |
| Consumption | 8 W |
|
| — |
General |
| |
| Storage temperature | From -20 to +70 ºC |
| Operating temperature | From -10 to +50 ºC |
| Internal temperature sensors | 2 sensors for reading internal temperature values on equipment |
| Humidity range | Up to 90 %RH non-condensing |
|
| — |
Housing |
| |
| Dimensions | 128 x 210 x 34 mm |
| Weight | 0.8 kg |
| Accessories | DIN Rail Connector |
| IP Code | IP50 |
|